Doppler Parameters Derived from Transthoracic Echocardiography Accurately Detect Bioprosthetic Mitral Valve Dysfunction.

Journal of the American Society of Echocardiography(2017)

引用 1|浏览42
暂无评分
摘要
•Transthoracic Doppler parameters identify dysfunctional mechanical mitral valves.•The authors examined these parameters in normal and abnormal bioprosthetic mitral valves.•The VTIMV/VTILVOT ratio was best at identifying valvular dysfunction.•VTIMV/VTILVOT > 2.5 was 100% sensitive for bioprosthetic mitral valve dysfunction.•VTIMV/VTILVOT > 2.5 was 95% specific for bioprosthetic mitral valve dysfunction.
更多
查看译文
关键词
Bioprosthetic mitral valve dysfunction,Echocardiography,Doppler,Transthoracic