Development of a two-dimensional imaging GEM detector using the resistive anode readout method with $6\times6$ cells

arXiv: Instrumentation and Detectors(2016)

引用 23|浏览16
暂无评分
摘要
We report the application of the resistive anode readout method on a two dimensional imaging GEM detector. The resistive anode consists of $6times6$ cells with the cell size $6~mathrm{mm}times6~mathrm{mm}$. New electronics and DAQ system are used to process the signals from the 49 readout channels. The detector has been tested by using the X-ray tube (8~keV). The spatial resolution of the detector is about $103.46~mathrm{{mu}m}$ with the detector response part about $66.41~mathrm{{mu}m}$. The nonlinearity of the detector is less than $1.5%$. A quite good two dimensional imaging capability is achieved as well. The performances of the detector show the prospect of the resistive anode readout method for the large readout area imaging detectors.