Novel discrimination parameters for neutron-gamma discrimination with liquid scintillation detectors using wavelet transform

JOURNAL OF INSTRUMENTATION(2015)

Cited 3|Views2
No score
Abstract
It has been observed that the discrimination performance of the wavelet transform method strongly depends on definition of discrimination parameters. These parameters are usually obtained from a combination of scaling functions at different scales, which represents the energy density of the wavelet coefficients. In this paper, the discrete wavelet transform (DWT) at minimum possible values of scale was investigated. Novel pulse shape discrimination parameters have been proposed for neutron and gamma discrimination in a mixed radiation field and tested with modeled pulses. The performance of these parameters was also validated in terms of quality of discrimination using experimental data of mixed events from an AmBe source collected with BC501 liquid scintillation detector. The quality of discrimination was evaluated by calculating a figure of merit (FOM) with all parameters under same experimental and simulation conditions. The FOM obtained with the proposed novel parameters was also compared with the charge comparison method. The proposed parameters exhibit better FOM as compared to the charge comparison method when high levels of noise are present in the data.
More
Translated text
Key words
Particle identification methods,Digital signal processing (DSP),Neutron detectors (cold, thermal, fast neutrons)