Investigation of balloon dilation devices for treatment of Eustachian tube dysfunction
Current Directions in Biomedical Engineering(2018)
摘要
AbstractThe presented three different balloon dilation devices for treating Eustachian tube dysfunction were compared regarding their geometries and designs. The balloon dilation behaviour was investigated by an in vitro test setup consisting of a test chamber with 37 °C water, a 2-axis laser scanner and a pressure controller. All balloons could be properly dilated up to the given nominal pressure (NP) or to rated burst pressure (RBP). The balloons reached their expected balloon diameter and length. The compliance data were 1.07 %/atm (TubaVent short), 1.16 %/atm (Acclarent Aera) and 1.28 %/atm (XprESS LoProfile). The measured profile and compliance data can be used for development of new devices for balloon dilation Eustachian tuboplasty.
更多查看译文
关键词
eustachian tube dysfunction,tube balloon dilation,balloon dilation eustachian tuboplasty (bet)