Evaluation of high resolution X-ray monochromatic wavelength dispersive X-ray fluorescence instrument for determining Pu and U in nuclear reprocessing streams

Kathryn McIntosh,George J. Havrilla, R.F. Gilmore, M.K. Holland

Spectrochimica Acta Part B: Atomic Spectroscopy(2018)

引用 4|浏览5
暂无评分
摘要
An alpha-prototype benchtop instrument was developed and tested for measuring Pu and U in spent fuel dissolver tanks in a nuclear fuel reprocessing plant. The high resolution X-ray (hiRX) instrument was designed with the aim of providing a new capability to measure samples with high radiation background with selectivity, accuracy, and precision using ultra-low volumes (7 μL) of liquid. The instrument was tested with specimens from the operational reprocessing facility at Savannah River National Laboratory. The results showed that bias was <6% for Pu nitrate process solution and < 18% for Pu in U/HNO3 matrix. Instrumental precision was <1% RSD while between microcell repeatability was <6% RSD. The limit of detection for Pu in HNO3 matrix was 3 × 10−5 g/L (100 s, 35 W). This study demonstrates that hiRX technology has the potential to provide rapid and simultaneous quantification of Pu and U without sample preparation, making it suitable for process monitoring and materials control and accountability applications.
更多
查看译文
关键词
MWDXRF,hiRX,Elemental analysis,Reprocessing,Accountancy