TOPAS Monte Carlo model of MD anderson scanning proton beam for simulation studies in proton therapy

BIOMEDICAL PHYSICS & ENGINEERING EXPRESS(2018)

引用 5|浏览16
暂无评分
摘要
A Monte Carlo model of the MD Anderson Cancer Center clinical scanning proton beam is created using the TOol for PArticle Simulations (TOPAS). This model is based on the commissioning data of integrated depth dose curves (IDDs) and lateral profiles of single proton spots. The mean range errors on the Bragg Peak (BP) for all energies are <0.1% of the range (mu = -0.01%, sigma = 0.05%). As an example of a possible application of this model, a quality assurance phantom measurement is simulated. The model can be used in further simulation studies of quality assurance, intensity modulated proton therapy, or MRI-guided proton therapy.
更多
查看译文
关键词
proton therapy,monte carlo simulations,beam model
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要