Development of a position-sensitive scintillation neutron detector for a new protein single-crystal diffractometer at J-PARC MLF

JOURNAL OF INSTRUMENTATION(2017)

引用 3|浏览19
暂无评分
摘要
A high-spatial-resolution, large-area position-sensitive scintillation- based neutron detector module has been developed for a new time-of-flight Laue single-crystal diffractometer to be constructed at J-PARC MLF. A first prototype detector implementing commercial Li-6: ZnS screens was produced based on a scintillator/wavelength-shifting fibre technology. The detector exhibited a spatial resolution of 2.5mm with a neutron-sensitive area of 320 x 320mm(2). We report on an initial evaluation of the detector performance, including its spatial resolution, detection efficiency and long-term background measurement, and also provide a brief description of a new neutron instrument.
更多
查看译文
关键词
Instrumentation and methods for time-of-flight (TOF) spectroscopy,Neutron detectors (cold, thermal, fast neutrons)