Stability assessment of radiation isocenter with the gimbaled linac system

International Journal of Cancer Therapy and Oncology(2017)

引用 0|浏览15
暂无评分
摘要
Purpose: We report the results of our year-long radiation isocenter accuracy verification for daily quality assurance (QA) implementation on a Vero4DRT system. Methods: The radiation isocenter was calculated using a cube phantom with a steel ball of diameter 10 mm fixed to the center of the phantom. A single photon beam was set with a field size of 100 × 100 mm 2 . Coincidence of the centroid of the steel ball at kiloVolt X-ray imaging isocenter and megaVolt beam radiation isocenter at each gantry and ring angle was tested. This procedure was performed for gantry angles of 0°, 90°, 180°, and 270°, and ring angles of 0°, 20°, and 340°. The centroid of the steel ball and the center of the radiation field were calculated to analyze the radiation isocenter error. This analysis was automatically calculated using the Daily Check tool in the Vero4DRT system. This QA was implemented between 24 August 2015 and 23 August 2016. Results: The average and standard deviation for pan and tilt directions were 0.12 ± 0.10 mm and -0.20 ± 0.13 mm, respectively. The maximum radiation isocenter accuracy error was 0.50 mm in both directions. Conclusion: The radiation isocenter alignment for the one year duration of the experiment was performed with high accuracy.
更多
查看译文
关键词
radiation isocenter,stability