Compensation scheme for online neutron detection using a Gd-covered CdZnTe sensor

Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment(2017)

引用 16|浏览4
暂无评分
摘要
The development of portable and personal neutron dosimeters requires compact and efficient radiation sensors. Gd-157, Gd-155 and Cd-113 nuclei present the highest cross-sections for thermal neutron capture among natural isotopes. In order to allow for the exploitation of the low and medium-energy radiative signature of the said captures, the contribution of gamma background radiation, falling into the same energy range, needs to be cancelled out. This paper introduces a thermal neutron detector based on a twin-dense semiconductor scheme. The neutron-sensitive channel takes the form of a Gd-covered CdZnTe crystal, a high density and effective atomic number detection medium. The background compensation will be carried out by means of an identical CdZnTe sensor with a Tb cover. The setting of a hypothesis test aims at discriminating the signal generated by the signature of thermal neutron captures in Gd from statistical fluctuations over the compensation of both independent channels. The measurement campaign conducted with an integrated single-channel chain and two metal Gd and Tb covers, under Cs-137 and Cf-252 irradiations, provides first quantitative results on gamma-rejection and neutron sensitivity. The described study of concept gives grounds for a portable, online-compatible device, operable in conventional to controlled environments.
更多
查看译文
关键词
Neutron detection,Compensation,Dense semiconductor,Gadolinium
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要