Detection of Mycobacterium chelonae, Mycobacterium abscessus Group, and Mycobacterium fortuitum Complex by a Multiplex Real-Time PCR Directly from Clinical Samples Using the BD MAX System

The Journal of Molecular Diagnostics(2017)

引用 10|浏览8
暂无评分
摘要
A new multiplex PCR test was designed to detect Mycobacterium chelonae, Mycobacterium abscessus group, and Mycobacterium fortuitum complex on the BD MAX System. A total of 197 clinical samples previously submitted for mycobacterial culture were tested using the new protocol. Samples were first treated with proteinase K, and then each sample was inoculated into the BD MAX Sample Buffer Tube. Extraction and multiplex PCR were performed by the BD MAX System, using the BD MAX ExK TNA-3 extraction kit and BD TNA Master Mix, along with specific in-house designed primers and probes for each target. The limit of detection of each target, as well as specificity, was evaluated. Of 197 clinical samples included in this study, 133 were positive and 60 were negative for mycobacteria by culture, and another 4 negative samples were spiked with M. chelonae ATCC 35752. The new multiplex PCR on the BD MAX had 97% concordant results with culture for M. abscessus group detection, 99% for M. chelonae, and 100% for M. fortuitum complex. The new multiplex PCR test performed on the BD MAX System proved to be a sensitive and specific test to detect M. chelonae, M. abscessus group, and M. fortuitum complex by real-time PCR on an automated sample-in results-out platform.