Development of a skull phantom for the assessment of implant X-ray visibility

Thomas Hoffmann, Fabian Klink,Axel Boese, Karin Fischer,Oliver Beuing,Georg Rose

Current Directions in Biomedical Engineering(2016)

引用 3|浏览4
暂无评分
摘要
Abstract The paper presents the development and test of a skull phantom, which can be used for the assessment of the radiographic visibility of neurovascular implants. State of the art methods are based on specimens of the human skull. These are highly individual and not suitable for comparison of different radiographic data sets. The development process of the skull phantom is described from data generation to image processing, design and manufacturing using 3D printing. An experimental setup is recommended to generate reproducible data sets for implant visibility assessment with bone mimicking structures of the phantom. The model is evaluated by qualitative comparison with equivalent data sets of the original human skull model. The results show, that contrast characteristics of the phantom and the human skull model are similar. X-ray attenuation of the human bone is higher than the polymeric phantom material. The introduced phantom allows the determination of X-ray attenuation characteristics of different neurovascular implants for medical approval and testing processes.
更多
查看译文
关键词
medical approval,phantom,radiopacity,skull,stent,testing,x-ray