Retrieval Of The Leadless Cardiac Pacemaker A Multicenter Experience

CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY(2016)

引用 47|浏览16
暂无评分
摘要
Background-Leadless cardiac pacemakers have emerged as a safe and effective alternative to conventional transvenous single-chamber ventricular pacemakers. Herein, we report a multicenter experience on the feasibility and safety of acute retrieval (<6 weeks) and chronic retrieval (>6 weeks) of the leadless cardiac pacemaker in humans.Methods and Results-This study included patients enrolled in 3 multicenter trials, who received a leadless cardiac pacemaker implant and who subsequently underwent a device removal attempt. The overall leadless pacemaker retrieval success rate was 94%: for patients whose leadless cardiac pacemaker had been implanted for <6 weeks (acute retrieval cohort), complete retrieval was achieved in 100% (n=5/5); for those implanted for >= 6 weeks (chronic retrieval cohort), retrieval was achieved in 91% (n=10/11) of patients. The mean duration of time from implant to retrieval attempt was 346 days (range, 88-1188 days) in the chronic retrieval cohort, and nearly two thirds (n=7; 63%) had been implanted for >6 months before the retrieval attempt. There were no procedure-related adverse events at 30 days post retrieval procedure.Conclusions-This multicenter experience demonstrated the feasibility and safety of retrieving a chronically implanted single-chamber (right ventricle) active fixation leadless pacemaker.
更多
查看译文
关键词
device removal,heart ventricles,humans,pacemaker,artificial,prostheses and implants