High-intensity focused ultrasound treatment in patients with refractory glaucoma

Graefes Archive for Clinical and Experimental Ophthalmology(2016)

Cited 42|Views7
No score
Abstract
Purpose To assess the safety and efficacy of ultrasound coagulation of the ciliary body in refractory glaucoma. Methods This prospective multicenter interventional study was conducted in two Italian university-affiliated glaucoma centers: St. Orsola-Malpighi Teaching Hospital (Bologna, Italy) and University Eye Clinic of Genoa (Genoa, Italy). The main inclusion criterion was the diagnosis of glaucoma with a baseline intraocular pressure (IOP) ≥ 21 mmHg while on maximum topical and systemic medical hypotensive treatment. The EyeOP1 device (Eye Tech Care, Rillieux-la-Pape, France), which was employed in the study, uses miniaturized transducers to produce high-intensity focused ultrasound (HIFU). Treatment consisted of the sequential activation of each transducer lasting 4 s (group 1), 6 s (group 2) or 8 s (group 3). Hypotensive medications were interrupted after surgery and then prescribed only if postoperative IOP was ≥ 21 mmHg during follow-up visits. Patients were assessed before and 1, 7, 14, 30, 90 and 180 days after the procedure. Primary outcomes were the mean IOP reduction in the overall population and in groups 1, 2 and 3, and the rates of complete success, qualified success and failure. Results Thirty eyes (16 open-angle, 10 angle-closure and 4 neovascular glaucoma) of 30 patients were included. The mean preoperative IOP was 30.1 ± 10.5 mmHg. Twenty-nine patients completed the entire study follow-up; one patient exited from the study 3 months after HIFU and underwent trabeculectomy. At days 1 and 180, the mean IOP was significantly reduced (18.4 ± 7.2 and 20.2 ± 6.2 mmHg, respectively; all p < 0.0001). Group 3 patients (8-s ultrasound exposure time) showed a greater IOP reduction than the other two groups (−16.2 ± 8.3 for group 3 vs. −8.8 ± 6.6 for group 2 and −3.7 ± 6.5 for group 1; p = 0.02 and p < 0.001, respectively). Qualified and complete success was achieved in 23.3 and 46.7% of patients, respectively; treatment failure was recorded in 6.6%. Conclusions Ultrasonic coagulation of the ciliary body is a safe and effective procedure for reducing IOP in refractory glaucoma. The increase in ultrasound exposure time appears to improve the response rate and the global efficacy of the procedure, with no detrimental effect on safety.
More
Translated text
Key words
Ultrasound coagulation,High-intensity focused ultrasound,Intraocular pressure reduction,Refractory glaucoma
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined