Comparison of the Wear Behaviour of Large Ceramic-on-Ceramic Bearings for Total Hip Arthroplasty
Journal of Bone and Joint Surgery-british Volume(2018)
Abstract
IntroductionDislocation is one of the major factors for revision surgery. Current literature states that the usage of larger bearing couples (> 36 mm) have the potential of reducing the risk of dislocation. Smaller ceramic-on-ceramic bearing couples (< 36 mm) have demonstrated very low wear rates. But does the wear behaviour change with increasing diameter? Therefore, the aim of this study was to compare wear rates of larger ceramic-on-ceramic bearing couples for total hip arthroplasty.Materials and MethodsWear tests according to ISO 14242 with 36, 40 and 44 mm zirconia platelet toughened alumina (ZPTA) bearings were performed in a servo-hydraulic hip simulator. In total, the specimens were loaded up to 5 million cycles. Wear was measured gravimetrically every million cycles. For each diameter three different combinations regarding clearance and roundness were chosen. One combination represented in tolerance parts (70 μm clearance, < 5 μm roundness). The other two combinations represented parts at the low...
MoreTranslated text
Key words
total hip arthroplasty,wear behaviour,total hip,ceramic-on-ceramic