A Prospective, Randomized Study of an Expanded Polytetrafluoroethylene Stent Graft versus Balloon Angioplasty for In-Stent Restenosis in Arteriovenous Grafts and Fistulae: Two-Year Results of the RESCUE Study

Journal of Vascular and Interventional Radiology(2016)

引用 48|浏览6
暂无评分
摘要
Abstract Purpose To assess the safety and efficacy of an expanded polytetrafluoroethylene stent graft versus balloon angioplasty for the treatment of in-stent restenosis in the venous outflow of hemodialysis access grafts and fistulae. Materials and Methods Two hundred seventy-five patients were randomized at 23 US sites to stent-graft placement or percutaneous transluminal angioplasty (PTA). Primary study endpoints were access circuit primary patency (ACPP) at 6 months and safety through 30 days; secondary endpoints were evaluated through 24 months. Results ACPP at 6 months was significantly higher in the stent-graft group (18.6%) versus the PTA group (4.5%; P P = .003 for noninferiority). The separation in ACPP survival curves remained through 12 months (stent graft, 6.2%; PTA, 1.5%). Treatment area primary patency (TAPP) was superior for the stent-graft group (66.4%) versus the PTA group (12.3%) at 6 months ( P P P Conclusions Stent-graft use provided better ACPP and TAPP than PTA when treating in-stent restenosis in patients receiving dialysis with arteriovenous grafts and fistulae.
更多
查看译文
关键词
ACPP,AV,CEC,CI,ePTFE,HR,IPF,ITT,KDOQI,PTA,TAPP