Calibration Of Soft X-Ray Detection With Synchrotron Radiation

HIGH ENERGY PHYSICS AND NUCLEAR PHYSICS-CHINESE EDITION(2004)

引用 23|浏览0
暂无评分
摘要
In this paper the calibrations of different soft X-ray detector elements used in experiment diagnostics of Inertial Confinement Fusion (ICF) with Synchrotron Radiation is reported. Beijing Synchrotron Radiation Facility used special operation type with beam current 35-110mA, the storage ring electron energy 2GeV, photon energy region 50-1500eV. The performance of the light source is studied by transmission grating and face array CCD system. We divide 50-1500eV energy region in seven energy section through different material prefilter with monochromaticity than 95 % The energy response of different soft X-ray detector elements are calibrated. The calibration results are used in ICF data analysis, which secures the accuracy of data treatment.