Micromachined Fused Silica Liquid Core Waveguide Capillary Flow Cell

Karsten Kraiczek, John Mannion, Susan Post, Andriy Tsupryk, Varun Raghunathan,Reid Brennen,Roland Zengerle

ANALYTICAL CHEMISTRY(2016)

引用 13|浏览4
暂无评分
摘要
A planar, chip-based flow cell for UV-vis absorbance detection in HPLC is presented. The device features a microfabricated free-standing liquid core waveguide (LCW) capillary detection tube of long path length that is based on total internal reflection. We report on the linearity and calibration slope characteristics of lithographically produced LCWs with different interior/exterior geometries. 3D ray tracing was indispensable in modeling behavior in the more demanding geometries: multipath behavior may be intrinsic to these waveguides with consequent nonlinearity. Fortunately, nonlinearity in lithographically easy-to-produce waveguide geometries (such as with a flat, concave exterior and a round interior) is not as detrimental as might be initially expected. Experimental performance is predictably affected by the attainable surface quality of the LCW and efficient and reproducible coupling of the input light into the LCW.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要