Improved error detection using a divided treatment plan in volume modulated arc therapy

Reports of Practical Oncology & Radiotherapy(2019)

引用 1|浏览24
暂无评分
摘要
Aim We sought to improve error detection ability during volume modulated arc therapy (VMAT) by dividing and evaluating the treatment plan. Background VMAT involves moving a beam source delivering radiation to tumor tissue through an arc, which significantly decreases treatment time. Treatment planning for VMAT involves many parameters. Quality assurance before treatment is a major focus of research. Materials and methods We used an established VMAT prostate treatment plan and divided it into 12°×30° sections. In all the sections, only image data that generated errors in one segment and those that were integrally acquired were evaluated by a gamma analysis. This was done with five different patient plans. Results The integrated image data resulting from errors in each section was 100% (tolerance 0.5mm/0.5%) in the gamma analysis result in all image data. Division of the treatment plans produced a shift in the mean value of each gamma analysis in the cranial, left, and ventral directions of 94.59%, 98.83%, 96.58%, and the discrimination ability improved. Conclusion The error discrimination ability was improved by dividing and verifying the portal imaging.