Investigations of blood ammonia analysis: Test matrices, storage, and stability.

Clinical biochemistry(2017)

引用 13|浏览16
暂无评分
摘要
An assessment of blood ammonia concentration is common medical practice in the evaluation of an individual with an unexplained mental status change or coma. The determination of a blood ammonia level is most commonly done using a glutamate dehydrogenase (GLDH)-based assay, although there are many potential sources of artifact and the literature is inconsistent regarding key preanalytic issues. Using a GLDH-based assay, we first investigated matrix effects using three anticoagulants: heparin, EDTA and oxalate. Heparin-anticoagulated plasma was substantially less precise than EDTA- and oxalate-anticoagulated plasma. Oxalate-anticoagulated plasma showed a greater baseline of apparent ammonia than either heparin- or EDTA-derived plasma, presumably due to interferants. We then evaluated the stability of EDTA-anticoagulated plasma for assessment of ammonia when stored at 4°C,-14°C or -70°C. There was a linear increase of ammonia with storage at both 4°C and -14°C. Plasma kept at -70°C for up to three weeks showed no change in measured ammonia relative to the baseline determination. This work clarifies preanalytic conditions for which a precise determination of ammonia can be accomplished using a GLDH-based assay.
更多
查看译文
关键词
Ammonia,Plasma,Hyperammonemia
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要