Rapid determination of trace copper in animal feed based on micro-plate colorimetric reaction and statistical partitioning correction.

FOOD CHEMISTRY(2017)

引用 7|浏览3
暂无评分
摘要
The objective of this study was to develop a micro-plate based colorimetric assay for rapid and high-throughput detection of copper in animal feed. Copper ion in animal feed was extracted by trichloroacetic acid solution and reduced to cuprous ion by hydroxylamine. The cuprous ion can chelate with 2,2'-bicinchoninic acid to form a Cu-BCA complex which was detected with high sensitivity by micro-plate reader at 354 nm. The whole assay procedure can be completed within 20 min. To eliminate matrix interference, a statistical partitioning correction approach was proposed, which makes the detection of copper in complex samples possible. The limit of detection was 0.035 mu g/mL and the detection range was 0.1-10 mu g/mL of copper in buffer solution. Actual sample analysis indicated that this colorimetric assay produced results consistent with atomic absorption spectrometry analysis. These results demonstrated that the developed assay can be used for rapid determination of copper in animal feed. (C) 2016 Elsevier Ltd. All rights reserved.
更多
查看译文
关键词
Copper,2,2 '-Bicinchoninic acid,Colorimetric assay,Animal feed