Mri-Guided Focused Ultrasound Robotic System For The Treatment Of Bone Cancer

INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY(2017)

引用 16|浏览2
暂无评分
摘要
Background A novel MRI-conditional robot was developed that navigates a focused ultrasound (FUS) transducer. With this robotic system the transducer can access bones. The intended application is pain palliation from bone cancer using thermal ablation using FUS.Methods The robotic system has four computer-controlled axes (three linear and one angular). The robotic system was manufactured using a digital manufacturing 3D printer, using acrylonitrile butadiene styrene (ABS) plastic. MRI-conditional optical encoders were used to accurately control the robotic system.Results The robotic system was successfully tested for MRI safety and compatibility, using fast-gradient pulse sequences and a liquid phantom. The robotic system has been tested for its functionality for creating discrete and multiple (overlapping) lesions in a gel phantom.Conclusions An MRI-conditional FUS robotic system was developed that has the potential to create thermal lesions with the intention of treating bone cancer for the purpose of pain palliation. Copyright (C) 2016 John Wiley & Sons, Ltd.
更多
查看译文
关键词
ultrasound, robot, MRI, bone, cancer, pain