Implant Vertical Fractures Provoked by Laboratory Procedures: A Finite Element Analysis Inspired from Clinical Cases.

Implant dentistry(2016)

引用 7|浏览9
暂无评分
摘要
PURPOSE:To investigate the causes for internal implant fractures, which is suggested to be one of the reasons for marginal bone loss. MATERIALS AND METHODS:From a 14-year database of 6051 implants, 10 single implant vertical fractures were identified and the abutments were all castable abutments. The abutments presented contamination and irregularities at the internal connecting areas. The hypothesis was that perfect fit was disturbed by laboratory polishing procedures, and finite element analysis (FEA) using overcorrected and undercorrected castable abutment models were created and tested against a perfect fit model. RESULTS:The results from the FEA presented that both overcorrected and undercorrected models presented nonuniform excessive plastic strain distribution in the neck portion of the implants where clinically an implant fracture was noted. CONCLUSIONS:The results suggested that laboratory procedures could induce plastic strain of the implant-abutment complex, which increases the risk of fracture.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper