Quantification of Iohexol in Serum by High-Performance Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS).

CLINICAL APPLICATIONS OF MASS SPECTROMETRY IN DRUG ANALYSIS: METHODS AND PROTOCOLS(2016)

引用 6|浏览11
暂无评分
摘要
Iohexol is a nonradioactive contrast medium, and its clearance from serum or urine is used to measure glomerular filtration rate (GFR). GFR is the most useful indicator of kidney function and progression of kidney disease. GFR determination using iohexol clearance is increasingly being applied in clinical practice, given its advantages over and correlation with inulin. We describe a high-performance liquid chromatography tandem mass spectrometry (LC-MS/MS) method for iohexol clearance, requiring only 50 μL of serum. The sample preparation involves protein precipitation with LC/MS-grade methanol, containing ioversol as the internal standard. Samples are centrifuged and supernatant is dried under nitrogen gas at room temperature. Samples are reconstituted with mobile phase (ammonium acetate-formic acid-water). Iohexol is separated using an HPLC gradient method on a C-8 analytical column. MS/MS detection is in the multiple-reaction monitoring (MRM) mode and the transitions monitored are m/z 822.0 to m/z 804.0 and m/z 807.0 to m/z 588.0 for iohexol and ioversol, respectively.
更多
查看译文
关键词
Glomerular filtration rate,Iohexol,Liquid chromatography,Mass spectrometry,Quantification,Serum