[Three-dimensional isotropic magnetic resonance imaging of the ankle joint].

Nan fang yi ke da xue xue bao = Journal of Southern Medical University(2015)

引用 0|浏览1
暂无评分
摘要
OBJECTIVE:To compare the image quality of isotropic 3-dimensional fast spin echo (3D-FSE), 3D fast field echo (3D-FFE), and 2D fast spin echo (2D-FSE) sequences in magnetic resonance imaging (MRI) of the anatomical structure of the ankle joint. METHODS:The ankle joints of 10 healthy volunteers were examined with isotropic 3D-FSE, 3D-FFE and 2D-FSE sequences using a 1.5T MR scanner and 3D reconstruction. The signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) of the tissues were measured. Two radiologists evaluated the image quality of the 3 sequences using a 5-point Likert scale in a double-blinded manner. RESULTS:The 3D-FSE sequences resulted in the highest SNRs for all the tissues and the highest CNRs for differentiation between cartilage and marrow, between muscle and tendon, and between tendon and fluid. In the estimation of image quality for cartilages, 3D-FFE had the highest score followed by 3D-FSE, and the latter had the highest score among the 3 sequences in displaying the tendon. CONCLUSION:3D-FSE sequence has a high performance in displaying the anatomical structures of complex joints especially for cartilage, ligament, and tendon tissues.
更多
查看译文
关键词
ankle,3-dimensional isotropic fast spin echo,3-dimensional fast field echo,two-dimensional fast spin echo,mag-netic resonance imaging