Investigation of combined kV/MV CBCT imaging with a high-DQE MV detector.

MEDICAL PHYSICS(2019)

引用 4|浏览14
暂无评分
摘要
PurposeCombined kV-MV cone-beam tomography (CBCT) imaging has been proposed for two potentially important image-guided radiotherapy applications: (a) scan time reduction (STR) and (b) metal artifact reduction (MAR). However, the feasibility of these techniques has been in question due to the low detective quantum efficiencies (DQEs) of commercially available electronic portal imagers (EPIDs). The goal of the work was to test whether a prototype high DQE MV detector can be used to generate acceptable quality pretreatment CBCT images at acceptable dose levels. Methods6MV and 100 kVp projection data were acquired on a Truebeam system (Varian, Palo Alto, CA). The MV data were acquired using a prototype EPID containing two scintillators (a) a standard copper-gadolinium oxysulfide (Cu-GOS) screen having a zero-frequency DQE (DQE(0)) value of 1.4%, and (b) a prototype-focused cadmium tungstate (CWO) pixelated strip with a DQE(0) = 22%. The kV data were acquired using the standard onboard imager (DQE(0) = 70%). The angular spacing of the MV projections was 0.81 degrees and the source output was 0.03 MU/projection while the kV projections were acquired with an angular spacing of 0.4 degrees at 0.3 mAs/projection. Image quality was evaluated using (a) an 18-cm diameter electron density phantom (CIRS, Norfolk, VA) with nine contrast inserts and (b) the resolution section of the 20-cm diameter Catphan phantom (The Phantom Laboratory, Greenwich, NY). For the MAR studies, two opposing CIRS phantom inserts were replaced by steel rods. The reconstruction methods were based on combining MV and kV data into one sinogram. The MAR reconstruction utilized mostly kV raw data with only those rays corrupted by metal requiring replacement with MV data (total absorbed dose = 0.7 cGy). For the STR study, projections from partially overlapping 105 degrees kV and MV acquisitions were combined to create a complete dataset that could have been acquired in 18 sec (absorbed dose = 2.5 cGy). MV-only (4.3 cGy) and kV-only (0.3 cGy) images were also reconstructed. ResultsThe average signal-to-noise ratio (SNR) of the inserts in the MV-only CWO and GOS CIRS phantom images were 0.62x and 0.12x the SNR of the inserts in kV-only image, respectively. The limiting spatial resolutions in the MV-only GOS, MV-only CWO, and kV-only Catphan images were 3, 6, and 8 lp/cm, respectively. In the combined kV/CWO STR reconstruction, all contrast inserts were visible while only two were detectable in the kV/Cu-GOS image due to high levels of noise (average SNRs of kV/CWO and kV/GOS inserts were 0.97x and 0.18x the SNR of the kV-only inserts, respectively). In the kV-MV MAR reconstructions, streaking artifacts were substantially reduced with all inserts becoming clearly visible in the kV/CWO image while only two were visible in the kV/Cu-GOS image (average SNRs of the kV/CWO and kV/Cu-GOS CIRS with metal inserts were 0.94x and 0.35x the SNRs of the kV-only CIRS without metal inserts). ConclusionsWe have demonstrated that a high-DQE MV detector can be applied to generating high-quality combined kV-MV images for SRT and MAR. Clinically acceptable doses were utilized.
更多
查看译文
关键词
CBCT imaging,high DQE,metal artifact reduction,MV detector,scan time reduction
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要