C3d Assay in Correlation With Single Antigen Bead Assay for Human Leukocyte Antigen Antibodies.
Transplantation Proceedings(2018)
Abstract
•The maximal MFI values of the SAB in the samples that were C3d assay-positive were significantly higher than those that were C3d assay-negative.•The MFI value of antibodies identified by SAB correlated with their positivity by C3d assay.•Only 0.4% of the antibodies with MFI <10,000 on SAB were identified by C3d assay with MFI >1000, whereas 59.4% of the antibodies with MFI >10,000 on SAB were identified by C3d assay with MFI >1000.
MoreTranslated text
Key words
human leukocyte antigen,single antigen bead,antibodies,assay