Estimating testosterone concentrations in adolescent girls: Comparison of two direct immunoassays to liquid chromatography-tandem mass spectrometry.

Steroids(2018)

引用 6|浏览10
暂无评分
摘要
CONTEXT:The appropriate role of direct total testosterone (T) immunoassays in reproductive research is controversial. OBJECTIVE:To assess the concordance between two direct immunoassays and a liquid chromatography-tandem mass spectrometry (LC-MS/MS) assay for total T in adolescent girls with measured concentrations < 50 ng/dl. DESIGN:Cross-sectional analysis. SETTING:Academic medical center. PARTICIPANTS:Adolescent girls (age 8.4-18.1 years) participating in clinical research protocols. INTERVENTION:Paired blood samples were obtained for total T by LC-MS/MS (n = 66; Mayo Clinic Laboratory) and by direct immunoassay (Center for Research in Reproduction)-either radioimmunoassay (RIA; n = 31) or chemiluminescence immunoassay (CLIA; n = 35). At the time of assay, laboratories were unaware that results would be compared. MAIN OUTCOME MEASURE:Measurement agreement between immunoassay and LC-MS/MS. RESULTS:Measured T concentrations (LC-MS/MS) were <7 to 44 ng/dl. The average difference between RIA and LC-MS/MS was 0.84 [-0.89, 2.56] ng/dl (mean [95% confidence interval]). RIA correlated very strongly with LC-MS/MS (r = 0.899; p < 0.0001); and both Deming regression and Bland-Altman analysis suggested no bias. The average difference between chemiluminescence and LC-MS/MS was 1.39 [-0.83, 3.60] ng/dl. CLIA correlated strongly with LC-MS/MS (r = 0.806; p < 0.0001). While Bland-Altman analysis suggested no systematic bias, Deming regression analysis suggested that, as measured values increased, values obtained by CLIA tended to be progressively, albeit only modestly, higher than those obtained by LC-MS/MS. CONCLUSIONS:These data support the use of rigorously-performed and carefully-validated direct T immunoassays in high-quality endocrine research in peripubertal adolescent girls.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要