Performance Evaluation of the Beckman Coulter DxN VERIS Hepatitis B Virus (HBV) Assay in Comparison With the Abbott RealTime HBV Assay.

ANNALS OF LABORATORY MEDICINE(2019)

Cited 5|Views22
No score
Abstract
The detection and quantification of hepatitis B virus (HBV) DNA plays an important role in diagnosing and monitoring HBV infection as well as in assessing the therapeutic response. We compared the analytical performance of a random access, fully automated HBV assay-DxN VERIS Molecular Diagnostics System (Beckman Coulter, Brea, CA, USA)-with that of Abbott RealTime HBV assay (Abbott Laboratories, Des Plaines, IL, USA). The between-day precision of the VERIS assay ranged from 0.92% (mean 4.68 log IU/mL) to 4.15% (mean 2.09 log IU/mL) for pooled sera from HBV patients. HBV DNA levels measured by the VERIS HBV assay correlated with the calculated HBV DNA levels (r(2) = 0.9994; P < 0.0001). The lower limit of quantification was estimated as 8.76 IU/mL (Probit analysis, 95% confidence interval: 7.32-12.00 IU/mL). Passing-Bablok regression analysis showed good concordance between the VERIS and RealTime assays for 187 chronic HBV samples (y = -0.2397+0.9712x; r = 0.981), as well as for 20 drug-resistant HBV genotype C positive samples (y = -0.5415+0.9954x; r = 0.961). The VERIS assay demonstrated performance similar to the RealTime assay and is suitable for high-throughput HBV DNA monitoring in large hospital laboratories.
More
Translated text
Key words
Analytical performance,HBV DNA monitoring,Beckman Coulter DxN VERIS HBV assay,Abbott RealTime HBV assay
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined