Myocardial Thermal Ablation with a Transesophageal High-Intensity Focused Ultrasound Probe: Experiments on Beating Heart Models.

Ultrasound in medicine & biology(2018)

引用 5|浏览13
暂无评分
摘要
Described here is a study of transesophageal thermal ablation of isolated and perfused beating hearts and non-human primates. An endoscope integrating a transesophageal echocardiography probe and a high-intensity focused ultrasound transducer was built and tested on five Langendorff-isolated hearts and three 30-kg baboons. B-Mode ultrasound, passive elastography and magnetic resonance imaging were performed to monitor thermal lesions. In isolated hearts, continuous and gated sonication parameters were evaluated with acoustic intensities of 9-12 W/cm2. Sonication parameters of gated exposures with 12 W/cm2 acoustic intensity for 5 min consistently produced visible lesions in the ventricles of isolated hearts. In animals, left atria and ventricles were exposed to repeated continuous sonications (4-15 times for 16 s) at an acoustic intensity at the surface of the transducer of 9 W/cm2. Clinical states of the baboons during and after the treatment were good. One suspected lesion in the left ventricle could be evidenced by elastography, but was not confirmed by magnetic resonance imaging. The transesophageal procedure therefore has the potential to create thermal lesions in beating hearts and its safety in clinical practice seems promising. However, further technical exploration of the energy deposition in the target would be necessary before the next pre-clinical experiments.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要