Long-term electrical performance of attain performa quadripolar left ventricular leads with all steroid-eluting electrodes: Results from a large worldwide clinical trial.

PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY(2018)

引用 6|浏览15
暂无评分
摘要
BackgroundSteroid-eluting (SE) electrodes suppress local inflammation and lower pacing capture thresholds (PCT); however, their effectiveness on quadripolar left ventricular (LV) leads in the cardiac vein is not fully studied. We evaluated the effectiveness of SE on all four LV pacing electrodes in human subjects enrolled in the Medtronic Attain (R) Performa quadripolar LV lead study. MethodsA total of 1,097 subjects were included in this evaluation. At each follow-up visit (1, 3, 6, and 12 months), LV PCT and pacing impedance were measured using either manual or automated testing methods. Summary statistics for PCT and impedance values were obtained for implant and each scheduled follow-up visit for all lead models. ResultsAverage extended bipolar (LV electrode to right ventricular Coil) PCTs for the four LV SE pacing electrodes (LV1, LV2, LV3, and LV4) on the three shapes of the quadripolar LV leads were 1.06 0.97 V, 1.38 +/- 1.26 V, 1.51 +/- 1.33 V, and 2.25 +/- 1.63 V, respectively, at 0.5-ms pulse width. PCTs remained low and stable throughout the 12-month follow-up period. ConclusionThis clinical trial demonstrated that SE on all LV pacing electrodes is associated with low and stable PCTs for all quadripolar LV lead electrodes, resulting in multiple viable vectors for LV pacing. The large number of available vectors facilitates basal pacing, avoidance of PNS, and potentially prolongs generator longevity due to lower PCTs.
更多
查看译文
关键词
cardiac resynchronization therapy defibrillator,LV lead,pacing capture threshold,quadripolar,steroid elution
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要