New AMH assay allows rapid point of care measurements of ovarian reserve.

GYNECOLOGICAL ENDOCRINOLOGY(2017)

引用 11|浏览4
暂无评分
摘要
In this study, we compare two commercial automated immunoassays used to evaluate serum anti-Mullerian hormone (AMH) levels as a prognostic value for ovarian response and pregnancy outcome in assisted reproductive technology cycles. Serum AMH was measured for 193 women. We performed a simultaneous measurement in serum AMH with the two alternative kits VIDAS (R) and Elecsys (R) AMH assay. For all women undergoing in vitro fertilization cycle, we collected data on their antral follicle count (AFC) and numbers of retrieved cumulus oocyte complexes (OC) and metaphase II oocytes and pregnancy outcome. The AMH values provided by VIDAS (R) were correlated with the values obtained with Elecsys (R) (0.977 for fresh and 0.971 for the frozen samples). For both assays AMH exhibited a moderate positive correlation with AFC, OC and MII oocytes (0.612, 0.674, 0.605 for VIDAS (R) and 0.570, 0.617, 0.530 for Elecsys (R), respectively). AMH prediction of biochemical and clinical pregnancy was similar. The present results suggest that the VIDAS (R) AMH assay is broadly comparable to the Elecsys-AMH assay in terms of technical performance for clinical or epidemiological use. Both automated assays performed in a similar way and the choice of assay can be made depending on the technical configuration of each laboratory.
更多
查看译文
关键词
Anti-Mullerian hormone,clinical outcome,comparison of method,Elecsys AMH assay,VIDAS AMH assay