Pedicle screw insertion accuracy in terms of breach and reposition using a new intraoperative cone beam computed tomography imaging technique and evaluation of the factors associated with these parameters of accuracy: a series of 695 screws.
European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society(2017)
Abstract
Use of intraoperative CBCT as 2D-fluoroscopy or coupled with a navigation system for pedicle screw insertion is accurate in terms of breach occurrence and reposition. However, these rates depend on the classification or grading system used. Use of a navigation system does not decrease the risk of breach significantly. And the risk of breach is higher in non-degenerative conditions (trauma, scoliosis, infection, and malignancy disease) than in degenerative diseases.
MoreTranslated text
Key words
Accuracy,Pedicle screw insertion,Cone beam computed tomography (CBCT),Breach detection,Screw reposition,Navigation system