Faster clinical response to the onset of adverse events: A wearable metacognitive attention aid for nurse triage of clinical alarms.

PLOS ONE(2018)

引用 18|浏览26
暂无评分
摘要
OBJECTIVE:This study evaluates the potential for improving patient safety by introducing a metacognitive attention aid that enables clinicians to more easily access and use existing alarm/alert information. It is hypothesized that this introduction will enable clinicians to easily triage alarm/alert events and quickly recognize emergent opportunities to adapt care delivery. The resulting faster response to clinically important alarms/alerts has the potential to prevent adverse events and reduce healthcare costs. MATERIALS AND METHODS:A randomized within-subjects single-factor clinical experiment was conducted in a high-fidelity 20-bed simulated acute care hospital unit. Sixteen registered nurses, four at a time, cared for five simulated patients each. A two-part highly realistic clinical scenario was used that included representative: tasking; information; and alarms/alerts. The treatment condition introduced an integrated wearable attention aid that leveraged metacognition methods from proven military systems. The primary metric was time for nurses to respond to important alarms/alerts. RESULTS:Use of the wearable attention aid resulted in a median relative within-subject improvement for individual nurses of 118% (W = 183, p = 0.006). The top quarter of relative improvement was 3,303% faster (mean; 17.76 minutes reduced to 1.33). For all unit sessions, there was an overall 148% median faster response time to important alarms (8.12 minutes reduced to 3.27; U = 2.401, p = 0.016), with 153% median improvement in consistency across nurses (F = 11.670, p = 0.001). DISCUSSION AND CONCLUSION:Existing device-centric alarm/alert notification solutions can require too much time and effort for nurses to access and understand. As a result, nurses may ignore alarms/alerts as they focus on other important work. There has been extensive research on reducing alarm frequency in healthcare. However, alarm safety remains a top problem. Empirical observations reported here highlight the potential of improving patient safety by supporting the meta-work of checking alarms.
更多
查看译文
关键词
wearable metacognitive attention aid,faster clinical response,nurse triage,adverse events,clinical response
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要