Abstract
This study compared 2-D shear wave elastography (SWE) and transient elastography (TE) for liver fibrosis staging in patients with chronic hepatitis B (CHB) infection using liver biopsy as the reference standard. Patients with CHB infection who underwent liver biopsy were consecutively included. After exclusions, 257 patients were analyzed. Two-dimensional SWE resulted in a significantly higher rate of reliable measurements (98.1%, 252/257) than TE (93.0%, 239/257) (p = 0.011). Liver stiffness measurements of the two examinations exhibited a strong correlation (r = 0.835, p < 0.001). In patients given a confirmed histologic diagnosis, Spearman's rank coefficients were 0.520 in stage F0 (p < 0.001), 0.684 in stage F1 (p < 0.001), 0.777 in stage F2 (p < 0.001), 0.672 in stage F3 (p < 0.001) and 0.755 in stage F4 (p < 0.001). There were no significant differences between the areas under the receiver operating characteristic (ROC) curves of 2-D SWE and TE for liver fibrosis staging (all p values > 0.05). Two-dimensional SWE had diagnostic accuracy comparable to that of TE for liver fibrosis staging. The measurements that the two techniques provide are not interchangeable.