Activation of hip prostheses in high energy radiotherapy and resultant dose to nearby tissue.

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS(2017)

引用 2|浏览12
暂无评分
摘要
High energy radiotherapy can produce contaminant neutrons through the photonuclear effect. Patients receiving external beam radiation therapy to the pelvis may have high-density hip prostheses. Metallic materials such as those in hip prostheses, often have high cross-sections for neutron interaction. In this study, Thackray (UK) prosthetic hips have been irradiated by 18MV radiotherapy beams to evaluate the additional dose to patients from the activation products. Hips were irradiated in- and out-of field at various distances from the beam isocenter to assess activation caused in-field by photo-activation, and neutron activation which occurs both in and out-of-field. NaI(Tl) scintillator detectors were used to measure the subsequent gamma-ray emissions and their half-lives. High sensitivity Mg, Cu, P doped LiF thermoluminescence dosimeter chips (TLD-100H) were used to measure the subsequent dose at the surface of a prosthesis over the 12h following an in-field irradiation of 10,000MU to a hip prosthesis located at the beam isocenter in a water phantom. Fe-53, Mn-56, and V-52 were identified within the hip following irradiation by radiotherapy beams. The dose measured at the surface of a prosthesis following irradiation in a water phantom was 0.20mGy over 12h. The dose at the surface of prostheses irradiated to 200MU was below the limit of detection (0.05mGy) of the TLD100H. Prosthetic hips are activated by incident photons and neutrons in high energy radiotherapy, however, the dose resulting from activation is very small.
更多
查看译文
关键词
high energy radiotherapy,photonuclear effect,prostheses activation
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要