谷歌浏览器插件
订阅小程序
在清言上使用

An Effective Temperature Compensation Approach For Ultrasonic Hydrogen Sensors

REVIEW OF SCIENTIFIC INSTRUMENTS(2018)

引用 4|浏览9
暂无评分
摘要
Hydrogen is a kind of promising clean energy resource with a wide application prospect, which will, however, cause a serious security issue upon the leakage of hydrogen gas. The measurement of its concentration is of great significance. In a traditional approach of ultrasonic hydrogen sensing, a temperature drift of 0.1 degrees C results in a concentration error of about 250 ppm, which is intolerable for trace amount of gas sensing. In order to eliminate the influence brought by temperature drift, we propose a feasible approach named as linear compensation algorithm, which utilizes the linear relationship between the pulse count and temperature to compensate for the pulse count error (AN) caused by temperature drift. Experimental results demonstrate that our proposed approach is capable of improving the measurement accuracy and can easily detect sub-100 ppm of hydrogen concentration under variable temperature conditions. Published by AIP Publishing.
更多
查看译文
关键词
effective temperature compensation approach,ultrasonic,hydrogen,sensors
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要