Causes Of Artificial Urinary Sphincter Failure And Strategies For Surgical Revision: Implications Of Device Component Survival
NEUROUROLOGY AND URODYNAMICS(2018)
Abstract
Pressure-regulating balloon (PRB) malfunction is a common cause of artificial urinary sphincter failure, but failure due to pump or occlusive cuff malfunction is rare. Thus, a large set of patients may undergo PRB-only revision to restore device function without urethral exposure.