Development and clinical application of a bioluminescence enzyme immunoassay for oxytocin.

LUMINESCENCE(2018)

Cited 1|Views12
No score
Abstract
The development of a highly sensitive analytical method for oxytocin could be useful in the diagnosis and treatment of autistic spectrum disorder. We previously developed a colorimetric enzyme immunoassay (EIA) for plasma oxytocin measurement. In this study, we developed a method to measure oxytocin concentrations using a higher sensitivity bioluminescent EIA. Biotinylated oxytocin bridged with five lysine residues was used in a competitive format. The standard curve range for oxytocin was 1.0 to 1000 pg/assay. In addition, there was good correlation between the colorimetric and bioluminescent immunoassays in terms of measured oxytocin concentration (r = 0.9665, n = 48). The bioluminescent EIA for plasma oxytocin was more rapid and provided higher sensitivity than the colorimetric immunoassay, making it suitable for clinical application.
More
Translated text
Key words
autistic spectrum disorder,bioluminescence enzyme immunoassay,biotin-labelled antigen,oxytocin,peptide hormone