Biomechanical performance of the Actifit ® scaffold is significantly improved by selection of irrigation fluid

ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY(2018)

Cited 0|Views12
No score
Abstract
Purpose Clinical reports on meniscal scaffolds seem promising, albeit relatively paucity exists regarding their biomechanical behavior. The aim of the study is to delineate the impact of differing suture materials and the type as well as the temperature of the irrigation fluid on the pull-out strength of a polyurethane meniscal scaffold (Actifit ® ). Materials and methods 128 specimens were utilized with horizontal sutures and uniaxial load-to-failure testing was performed. We compared two different suture materials—polydioxanone (PDS) and non-absorbable, braided polyester sutures (NABP)—as well as two common irrigation fluids—lactated Ringer’s and electrolyte-free, hypotonic Mannitol-Sorbitol. All specimens were further evaluated according to two different temperatures [room temperature (20 °C) and near-core body temperature (37 °C)]. Results Mean load-to-failure was 53.3 ± 6.5 N. There was no significant difference between the NABP and the PDS group. Ringer group showed a significantly higher load-to-failure compared to Purisole ( P = .0002). This was equivalent for both PDS ( P = .0008) and NABP sutures ( P = .0008). Significantly higher failure loads could be established for the 37° group ( P = .041); yet, this difference was neither confirmed for the PDS or in in the NABP subgroup. Only the subgroup using Purisole at 37° showed significantly higher failure loads compared to 20° ( P = .017). Conclusions This study underlines the potential to improve pull-out strength during implantation of an Actifit ® scaffold by alteration of the type of irrigation fluid. Lactated Ringer solution provided the highest construct stability in regard to load-to-failure testing and should be considered whenever implantation of a polyurethane meniscal scaffold is conducted.
More
Translated text
Key words
Knee, Meniscus, Meniscal scaffold, Actifit (R), Biomechanical testing, Polyurethane
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined