Chrome Extension
WeChat Mini Program
Use on ChatGLM

The design and fabrication of beta sensor system for in situ diffusion tests in mudstone in France.

Applied Radiation and Isotopes(2017)

Cited 6|Views9
No score
Abstract
Long term in situ diffusion experiments in the Callovian-Oxfordian mudstone (France) are designed in the context of nuclear waste management. β-emitters HTO and 36Cl radiotracers are representative of neutral and anionic species in these experiments. Studies have been carried out to design an in situ beta monitoring system in order to quantify tracer migration in the rock pore water over time. The development, design and calibration of β-emitter radiation sensors were performed. An optimised geometry was calculated for the detection chamber of 36Cl in solution (Ø×L=30×2mm2) via Monte-Carlo N-Particle transport simulation tools. A better SNR (Signal to Noise Ratio) was obtained with YSO cylindrical crystal (Ø×L=6.5×0.5mm2). A sensitivity of 0.21cpm/Bq/mL was measured using 36Cl standards. The 36Cl detection limits were 18Bq/mL and 14.2Bq/mL after 1h and 24h of accumulation time respectively. Such beta sensors, placed 350mm from the radiotracer injection borehole, could successfully discriminate 36Cl anisotropic diffusion in various directions within 4 years. It will be tested in situ and can be adapted to other in situ experience.
More
Translated text
Key words
Mudstone,In situ,Radiotracer diffusion,Beta sensor,MCNP simulations
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined