Upgrade of the Mirnov probe arrays on the J-TEXT tokamak.

REVIEW OF SCIENTIFIC INSTRUMENTS(2017)

引用 21|浏览27
暂无评分
摘要
The magnetic diagnostic of Mirnov probe arrays has been upgraded on the J-TEXT tokamak to measure the magnetohydrodynamic instabilities with higher spatial resolution and better amplitude-frequency characteristics. The upgraded Mirnov probe array contains one poloidal array with 48 probe modules and two toroidal arrays with 25 probe modules. Each probe module contains two probes which measure both the poloidal and the radial magnetic fields (B-p and B-r). To ensure that the Mirnov probe possess better amplitude-frequency characteristics, a novel kind of Mirnov probe made of low temperature co-fired ceramics is utilized. The parameters and frequency response of the probe are measured and can meet the experiment requirement. The new Mirnov arrays have been normally applied for a round of experiments, including the observation of tearing modes and their coupling as well as high frequency magnetic perturbation due to the Alfven eigenmode. In order to extract useful information from raw signals, visualization processing methods based on singular value decomposition and cross-power spectrum are applied to decompose the coupled modes and to determine the mode number. Published by AIP Publishing.