Design and Evaluation of a Closed-Loop Anesthesia System With Robust Control and Safety System.

ANESTHESIA AND ANALGESIA(2018)

引用 48|浏览36
暂无评分
摘要
BACKGROUND: Closed-loop control of anesthesia involves continual adjustment of drug infusion rates according to measured clinical effect. The NeuroSENSE monitor provides an electroenceph-alographic measure of depth of hypnosis (wavelet-based anesthetic value for central nervous system monitoring [WAV(CNS)]). It has previously been used as feedback for closed-loop control of propofol, in a system designed using robust control engineering principles, which implements features specifically designed to ensure patient safety. Closed-loop control of a second drug, remifentanil, may be added to improve WAV(CNS) stability in the presence of variable surgical stimulation. The objective of this study was to design and evaluate the feasibility of a closed-loop system for robust control of propofol and remifentanil infusions using WAV(CNS) feedback, with an infusion safety system based on the known pharmacological characteristics of these 2 drugs. METHODS: With Health Canada authorization, research ethics board approval, and informed consent, American Society of Anesthesiologists I-III adults, requiring general anesthesia for elective surgery, were enrolled in a 2-phase study. In both phases, infusion of propofol was controlled in closed loop during induction and maintenance of anesthesia, using WAV(CNS) feedback, but bounded by upper- and lower-estimated effect-site concentration limits. In phase I, remifentanil was administered using an adjustable target-controlled infusion and a controller was designed based on the collected data. In phase II, remifentanil was automatically titrated to counteract rapid increases in WAV(CNS). RESULTS: Data were analyzed for 127 patients, of median (range) age 64 (22-86) years, undergoing surgical procedures lasting 105 (9-348) minutes, with 52 participating in phase I and 75 in phase II. The overall control performance indicator, global score, was a median (interquartile range) 18.3 (14.2-27.7) in phase I and 14.6 (11.6-20.7) in phase II (median difference, -3.25; 95% confidence interval, -6.35 to -0.52). The WAVCNS was within +/- 10 of the setpoint for 84.3% (76.6-90.6) of the maintenance of anesthesia in phase I and 88.2% (83.1-93.4) in phase II (median difference, 3.7; 95% confidence interval, 0.1-6.9). The lower propofol safety bound was activated during 30 of 52 (58%) cases in phase I and 51 of 75 (68%) cases in phase II. CONCLUSIONS: Adding closed-loop control of remifentanil improved overall controller performance. This controller design offers a robust method to optimize the control of 2 drugs using a single sensor. The infusion safety system is an important component of a robust automated anesthesia system, but further research is required to determine the optimal constraints for these safe conditions.
更多
查看译文
关键词
Consciousness Monitoring,Endoscopic Sedation
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要