Design and Evaluation of a Closed-Loop Anesthesia System With Robust Control and Safety System.
ANESTHESIA AND ANALGESIA(2018)
摘要
BACKGROUND: Closed-loop control of anesthesia involves continual adjustment of drug infusion rates according to measured clinical effect. The NeuroSENSE monitor provides an electroenceph-alographic measure of depth of hypnosis (wavelet-based anesthetic value for central nervous system monitoring [WAV(CNS)]). It has previously been used as feedback for closed-loop control of propofol, in a system designed using robust control engineering principles, which implements features specifically designed to ensure patient safety. Closed-loop control of a second drug, remifentanil, may be added to improve WAV(CNS) stability in the presence of variable surgical stimulation. The objective of this study was to design and evaluate the feasibility of a closed-loop system for robust control of propofol and remifentanil infusions using WAV(CNS) feedback, with an infusion safety system based on the known pharmacological characteristics of these 2 drugs. METHODS: With Health Canada authorization, research ethics board approval, and informed consent, American Society of Anesthesiologists I-III adults, requiring general anesthesia for elective surgery, were enrolled in a 2-phase study. In both phases, infusion of propofol was controlled in closed loop during induction and maintenance of anesthesia, using WAV(CNS) feedback, but bounded by upper- and lower-estimated effect-site concentration limits. In phase I, remifentanil was administered using an adjustable target-controlled infusion and a controller was designed based on the collected data. In phase II, remifentanil was automatically titrated to counteract rapid increases in WAV(CNS). RESULTS: Data were analyzed for 127 patients, of median (range) age 64 (22-86) years, undergoing surgical procedures lasting 105 (9-348) minutes, with 52 participating in phase I and 75 in phase II. The overall control performance indicator, global score, was a median (interquartile range) 18.3 (14.2-27.7) in phase I and 14.6 (11.6-20.7) in phase II (median difference, -3.25; 95% confidence interval, -6.35 to -0.52). The WAVCNS was within +/- 10 of the setpoint for 84.3% (76.6-90.6) of the maintenance of anesthesia in phase I and 88.2% (83.1-93.4) in phase II (median difference, 3.7; 95% confidence interval, 0.1-6.9). The lower propofol safety bound was activated during 30 of 52 (58%) cases in phase I and 51 of 75 (68%) cases in phase II. CONCLUSIONS: Adding closed-loop control of remifentanil improved overall controller performance. This controller design offers a robust method to optimize the control of 2 drugs using a single sensor. The infusion safety system is an important component of a robust automated anesthesia system, but further research is required to determine the optimal constraints for these safe conditions.
更多查看译文
关键词
Consciousness Monitoring,Endoscopic Sedation
AI 理解论文
溯源树
样例
![](https://originalfileserver.aminer.cn/sys/aminer/pubs/mrt_preview.jpeg)
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要