Real-time detection system for tumor localization during minimally invasive surgery for gastric and colon cancer removal: In vivo feasibility study in a swine model.

JOURNAL OF SURGICAL ONCOLOGY(2018)

引用 11|浏览2
暂无评分
摘要
Background and ObjectivesDuring minimally invasive surgery (MIS), it is impossible to directly detect marked clips around tumors via palpation. Therefore, we developed a novel method and device using Radio Frequency IDentification (RFID) technology to detect the position of clips during minimally invasive gastrectomy or colectomy. MethodsThe feasibility of the RFID-based detection system was evaluated in an animal experiment consisting of seven swine. The primary outcome was to successfully detect the location of RFID clips in the stomach and colon. The secondary outcome measures were to detect time (time during the intracorporeal detection of the RFID clip), and accuracy (distance between the RFID clip and the detected site). ResultsA total of 25 detection attempts (14 in the stomach and 11 in the colon) using the RFID antenna had a 100% success rate. The median detection time was 32.5s (range, 15-119s) for the stomach and 28.0s (range, 8-87s) for the colon. The median detection distance was 6.5mm (range, 4-18mm) for the stomach and 6.0mm (range, 3-13mm) for the colon. ConclusionsWe demonstrated favorable results for a RFID system that detects the position of gastric and colon tumors in real-time during MIS.
更多
查看译文
关键词
early colon cancer,early gastric cancer,intraoperative tumor localization,minimally invasive surgery,radio frequency identification detector
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper