Assessment of micro-dose biplanar radiography in lower limb measurements in children

European Radiology(2017)

Cited 1|Views20
No score
Abstract
Objectives To evaluate in children microdose protocol compared with low dose for lower limb alignment (LLA) measurements on biplanar radiography. Methods Children 6 years or older were included. Height, weight and hip width were measured prior to imaging. Hip width allowed classifying children into morphotype M1 (width <25 cm), M2 (25–35 cm) or M3 (>35 cm) corresponding to predefined acquisition parameters (kV, mA, tube speed). Micro- and low-dose protocols were used alternately, with simultaneous acquisition of frontal and lateral radiographs. LLA measurements were performed by two independent observers (n = 526). In 15 children per morphotype, a third observer performed measurements twice (n = 180). Intraclass correlation coefficients and the dose (delivered, absorbed) were calculated. Results 100 girls and 160 boys (mean age = 11.7 years) were investigated: 74 M1 (mean BMI, 15.7kg/m 2 ), 149 M2 (19.8 kg/m 2 ) and 40 M3 (30.2 kg/m 2 ). With microdose, inter- and intra-observer agreement was >0.90 for lengths whatever the morphotype, 0.75–0.90 (M1) and >0.90 (M2, M3) for valgus/varus and flexion/hyperextension deviations. Dose reduction reached a factor of 8.5 and 5.4 for the delivered and absorbed dose respectively. Conclusions Microdose could be used for LLA measurements in children and permits a significant dose reduction. Key points • Lower limb lengths of children can be evaluated with microdose biplanar radiography. • Valgus/varus deviations also can be evaluated with microdose biplanar radiography. • Microdose biplanar radiography significantly reduces delivered and absorbed dose in children.
More
Translated text
Key words
Child,Lower limbs,Biplanar radiographs,Microdose,Radiation dose
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined