Development and Validation of Stereo Selective Method for the Separation of Razoxane Enantiomers in Hydrophilic Interaction Chromatography.

JOURNAL OF CHROMATOGRAPHIC SCIENCE(2018)

Cited 7|Views1
No score
Abstract
A novel sensitive and high throughput chiral hydrophilic interaction chromatographic (HILIC) method was developed to separate razoxane enantiomers namely levrazoxane (R-isomer) and dexrazoxane (DEX) in pharmaceutical active ingredient samples. A systematic chiral chromatographic screening system was employed in using multiple HPLC chromatographic modes on various polysaccharide based chiral columns to obtain a potential separation between enantiomers. HPLC separation was achieved using a mobile phase of aqueous 10 mM ammonium bicarbonate and mixture of organic modifiers (70/30, v/v) in the ratio of (5/95, v/v) on an immobilized polysaccharide based chiral stationary phase namely CHIRALPAK IE-3. The chromatographic resolution between the enantiomers was found to be not < 8 in the developed method. The values of the limit of detection and limit of quantification of DEX and levrazoxane were found to be 0.0037, 0.011 and 0.0043, 0.013 mu gmL(-1), respectively. The validated method yielded good results regarding precision, linearity, selectivity and found to be superior in sensitivity when compared to reported method for the accurate quantification of undesired enantiomer.