CHARACTERIZATION OF FLUORESCENT NUCLEAR TRACK DETECTORS AS CRITICALITY DOSIMETERS II.

RADIATION PROTECTION DOSIMETRY(2018)

Cited 6|Views9
No score
Abstract
Fluorescent nuclear track detectors (FNTDs) as criticality dosimeters for both neutrons and gamma are further characterized in terms of angular dependence and quick dose assessment. The power spectrum integral depth profiles obtained from stacks of fluorescent images acquired within FNTDs exposed to a broad spectrum neutron field at various angles are analyzed to determine a calibration curve for angular dependence. MCNPX simulations were shown to be in good agreement with experimental results. A prototype triage reader was also designed and tested for quick assessment of dose. An unfolding technique incorporating both energy dependence and angular dependence is discussed. The advantages and shortcomings of using FNTDs in the event of a criticality excursion accident are analyzed.