Detection and Quantification of Inorganic and Organic Anions in Natural, Potable, and Wastewaters in Northern New York Using Capillary Zone Electrophoresis and Indirect UV Detection.

Journal of chromatography & separation techniques(2017)

引用 3|浏览2
暂无评分
摘要
Capillary zone electrophoresis (CZE) is a sensitive and rapid technique used for determining traces of inorganic and organic anions in potable, natural, and wastewaters. Here, CZE with indirect UV-diode array detection (CZE-DAD) was employed with a background electrolyte system comprising of an Agilent Technologies proprietary basic anion buffer at pH 12.0 and a forensic anion detection method. The limits of detection (LOD) for this method ranged between 3 and 5 ppm and involved hydrodynamic injection of 50 mbar for 6 s with a negative polarity separation voltage of -30 kV at 30°C, a detection wavelength of 350 nm and indirect reference of 275 nm. Fourteen different anions were checked for in the water samples that were examined and included bromide, chloride, thiosulfate, nitrate, nitrite, sulfate, azide, carbonate, fluoride, arsenate, phosphate, acetate, lactate, and silicate. The water samples were collected from Northern New York towns and the Raquette River water system, the third longest river in New York State and the largest watershed of the central and western Adirondacks. The concentrations detected for these anions ranged from <5.0 ppm to 260 ppm.
更多
查看译文
关键词
Adirondack watershed,Capillary zone electrophoresis,Indirect UV detection,Northern New York Raquette River,Organic and inorganic anions