Molecular absorption measurements with an optical fibre coupled array of ultra-violet light-emitting diodes.

Analytica Chimica Acta(2017)

引用 7|浏览6
暂无评分
摘要
A photometric detector based on eight different light-emitting diodes covering the ultraviolet range from 255 nm to 350 nm is described. These are coupled with fused silica optical fibres to a conventional cuvette with 1 cm optical path length or to a low volume flow through cell for detection in high-performance liquid chromatography. Photodiodes are employed for the measurement of the transmitted intensity as well as of a reference signal and the photocurrents are processed with a log-ratio amplifier to obtain a voltage proportional to absorbance values. The wavelength desired for the measurement at hand is selected by electronically switching on the requisite light-emitting diode. The detector was found to have a low noise level of 80 μAU. In batch-wise measurements as well as in detection for high-performance liquid chromatography dynamic ranges of 2–3 orders of magnitude were possible. Reproducibilities in peak areas for the latter application were better than 1%.
更多
查看译文
关键词
Light-emitting diode,Ultra-violet,Array,Photodiode,Photometry,Liquid chromatography