Maximizing Ion Transmission in Differential Mobility Spectrometry

Journal of The American Society for Mass Spectrometry(2017)

Cited 11|Views0
No score
Abstract
We provide modeling and experimental data describing the dominant ion-loss mechanisms for differential mobility spectrometry (DMS). Ion motion is considered from the inlet region of the mobility analyzer to the DMS exit, and losses resulting from diffusion to electrode surfaces, insufficient effective gap, ion fragmentation, and fringing field effects are considered for a commercial DMS system with 1-mm gap height. It is shown that losses due to diffusion and radial oscillations can be minimized with careful consideration of residence time, electrode spacing, gas flow rate, and waveform frequency. Fragmentation effects can be minimized by limitation of the separation field. When these parameters were optimized, fringing field effects at the DMS inlet contributed the most to signal reduction. We also describe a new DMS cell configuration that improves the gas dynamics at the mobility cell inlet. The new cell provides a gas jet that decreases the residence time for ions within the fringing field region, resulting in at least twofold increase in ion signal as determined by experimental data and simulations. Graphical Abstract ᅟ