Glyoxal measurement with a proton transfer reaction time of flight mass spectrometer (PTR-TOF-MS): characterization and calibration.

JOURNAL OF MASS SPECTROMETRY(2017)

引用 29|浏览10
暂无评分
摘要
We examine the potential for PTR-TOF-MS systems to quantitatively measure glyoxal in ambient air by characterizing the response of the instrument to a dilute glyoxal sample, calibrating the system as a function of humidity. The concentration of glyoxal in a sample air-stream was measured with an UV absorption spectrometer in parallel to a PTR-TOF-MS. This calibration demonstrated that the PTR-TOF-MS has a relatively low sensitivity to glyoxal particularly at high humidity. Extensive fragmentation of glyoxal to formaldehyde was observed. This behaviour not only desensitizes PTR-MS system to glyoxal; it may also pose a problem to the quantification of formaldehyde. (C) 2016 The Authors. Journal of Mass Spectrometry Published by John Wiley & Sons Ltd.
更多
查看译文
关键词
PTR-TOF-MS,glyoxal,fragmentation,formaldehyde,glyoxal calibration
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要